工作原理:
1、手柄位進樣(Load)位置時,樣品經微量進樣針從進樣孔注射進定量環,定量環充滿後,多餘樣品從放空孔排出;
2、將手柄轉動至進樣(Inject)位置時,閥與液相流路接通,由泵輸送的流動相衝洗定量環,推動樣品進入液相分析柱進行分析。
雖然六通閥進樣器具有結構簡單、使用方便、壽命長、日(ri)常(chang)無(wu)需(xu)維(wei)修(xiu)等(deng)特(te)點(dian),但(dan)正(zheng)確(que)的(de)使(shi)用(yong)和(he)維(wei)護(hu)將(jiang)能(neng)增(zeng)加(jia)使(shi)用(yong)壽(shou)命(ming),保(bao)護(hu)周(zhou)邊(bian)設(she)備(bei),同(tong)時(shi)增(zeng)加(jia)分(fen)析(xi)準(zhun)確(que)度(du)。如(ru)使(shi)用(yong)得(de)當(dang)的(de)話(hua),六(liu)通(tong)閥(fa)進(jin)樣(yang)器(qi)一(yi)般(ban)可(ke)連(lian)續(xu)進(jin)樣(yang)3萬次而無需維修。
以下淺談有關六通閥進樣器的使用及保養事宜(僅供參考):
1、手柄處於Load和Inject之zhi間jian時shi,由you於yu暫zan時shi堵du住zhu了le流liu路lu,流liu路lu中zhong壓ya力li驟zhou增zeng,再zai轉zhuan到dao進jin樣yang位wei,過guo高gao的de壓ya力li在zai柱zhu頭tou上shang引yin起qi損sun壞huai,所suo以yi應ying盡jin快kuai轉zhuan動dong閥fa,不bu能neng停ting留liu在zai中zhong途tu。在zaiHPLCxitongzhongshiyongdezhusheqizhentouyoubieyuqixiangsepu,shipingtouzhusheqi。yifangmian,zhentouwaicejintiejinyangqimifengguanneice,mifengxingnenghao,bulouye,buyinrukongqi;另一方麵,也防止了針頭刺壞密封組件及定子。
2、六通閥進樣器的進樣方式有部分裝液法和完全裝液法兩種。使用部分裝液法進樣時,進樣量最多為定量環體積的75%,如20gL的定量環最多進樣15p,L的樣品,並且要求每次進樣體積準確、相同;使用完全裝液法進樣時,進樣量最少為定量環體積的3至5倍,即20gL的定量環最少進樣60至1001aL的樣品,這樣才能完全置換樣品定量環內殘留的溶液,達到所要求的精密度及重現性。推薦采用100ul的平頭進樣針配合20ul滿環進樣。
3、可根據進樣體積的需要自已製作定量環,一般不要求精確計算定量環的體積,譬如,一根名義上10gL的定量環,實際是9gL還是1lgL並不重要,因為被測樣品和校正樣品的進樣體積保持一致,在計算結果時誤差都被抵消了。
4、進樣樣品要求無微粒和能阻死針頭及進樣閥的物質,樣品溶液均要用0.45~tm的(de)濾(lv)膜(mo)過(guo)濾(lv)。防(fang)止(zhi)微(wei)粒(li)阻(zu)塞(sai)進(jin)樣(yang)閥(fa)和(he)減(jian)少(shao)對(dui)進(jin)樣(yang)閥(fa)的(de)磨(mo)損(sun)。為(wei)防(fang)止(zhi)緩(huan)衝(chong)鹽(yan)和(he)其(qi)它(ta)殘(can)留(liu)物(wu)質(zhi)留(liu)在(zai)進(jin)樣(yang)係(xi)統(tong)中(zhong),每(mei)次(ci)結(jie)束(shu)後(hou)應(ying)衝(chong)洗(xi)進(jin)樣(yang)器(qi),通(tong)常(chang)用(yong)不(bu)含(han)鹽(yan)的(de)稀(xi)釋(shi)劑(ji)、水或不含鹽的流動相衝洗,在進樣閥的Load和Inject位置反複衝洗,再用無纖維紙擦淨注射器針頭的外側。