1.對(dui)新(xin)購(gou)儀(yi)器(qi),電(dian)源(yuan)接(jie)通(tong),開(kai)關(guan)打(da)到(dao)加(jia)熱(re)位(wei)置(zhi),從(cong)顯(xian)微(wei)鏡(jing)中(zhong)觀(guan)察(cha)熱(re)台(tai)中(zhong)心(xin)光(guang)孔(kong)是(shi)否(fou)處(chu)於(yu)視(shi)場(chang)中(zhong),若(ruo)左(zuo)右(you)偏(pian),可(ke)左(zuo)右(you)調(tiao)節(jie)顯(xian)微(wei)鏡(jing)來(lai)解(jie)決(jue)。前(qian)後(hou)不(bu)居(ju)中(zhong),可(ke)以(yi)鬆(song)動(dong)熱(re)台(tai)兩(liang)旁(pang)的(de)兩(liang)隻(zhi)螺(luo)釘(ding),注(zhu)意(yi)不(bu)要(yao)拿(na)下(xia)來(lai),隻(zhi)要(yao)鬆(song)動(dong)就(jiu)可(ke)以(yi)了(le),然(ran)後(hou)前(qian)後(hou)推(tui)動(dong)熱(re)台(tai)上(shang)下(xia)居(ju)中(zhong)即(ji)可(ke),鎖(suo)緊(jin)兩(liang)隻(zhi)螺(luo)釘(ding)。在(zai)做(zuo)推(tui)動(dong)熱(re)台(tai)時(shi),為(wei)了(le)防(fang)止(zhi)熱(re)台(tai)燙(tang)傷(shang)手(shou)指(zhi),把(ba)波(bo)段(duan)開(kai)關(guan)和(he)電(dian)位(wei)器(qi)扳(ban)到(dao)編(bian)號(hao)最(zui)小(xiao)位(wei)置(zhi),即(ji)逆(ni)時(shi)針(zhen)旋(xuan)到(dao)底(di)。
2.進行升溫速率調整,這可用秒表式手表來調整。在秒表某一值時,記錄下這時的溫度值,然後,秒表轉一圈(一分鍾)時再記錄下溫度值。這樣連續記錄下去,直到你所要求測量的熔點值時,其升溫速率為1℃/分。太快或太慢可通過粗調和微調旋鈕來調節。注意即使粗調和微調旋鈕不動,但隨著溫度的升高,其升溫速率會變慢。
3.測溫儀的傳感器上,把其插入熱台孔到底即可,若其位置不對,將影響測量準確度。
4.要得到準確的熔點值,先用熔點標準物質進行測量標定。求出修正值。(修正值=標準值-所測熔點值),作為測量時的修正依據。注意:標準樣品的熔點值應和你所需的樣品熔點值越接近越好。這時,樣品的熔點值=該樣品實測值+修正值)。
5.對待測樣品要進行幹燥處理,或放在幹燥缸內進行幹燥,粉末要進行研細。
6.當采用載—蓋玻片測量時,建議將蓋玻片(薄的一塊)放在熱台上,放上藥粉,再放上載玻片測量。
7.在數字溫度顯示最小一位(如8和7之間跳動時)應讀為8.5℃。
8.在重複測量時,開關處於中間關的狀態,這時加熱停止。自然冷卻到10℃以下時,放入樣品,開關打到加熱時,即可進行重複測量。
9.測試完畢,應切斷電源,當熱台冷卻到室溫時,方可將儀器裝入包箱內。